National Stock Number:
6530-01-487-6868

Federal Supply Class:
6530

National Item Identification Number:
014876868

Description:
THERMOREGULATOR,PATIENT

Detail:
AN ITEM DESIGNED TO LOWER OR RAISE THE BODY TEMPERATURE OF THE PATIENT DURING SURGERY, AS WELL AS QUICKLY REDUCING AND CONTROLLING DANGEROUSLY HIGH TEMPERATURES OF POSTOPERATIVE INFECTIONS. TEMPERATURE READINGS CAN BE MADE ESOPHAGEALLY OR RECTALLY BY THE USE OF THERMISTOR SENSING PROBES.


Manufacturer Information:
01-035564616DEFENSE MEDICAL STANDARDIZATION BOARD


Techincal Specification:
GENERAL CHARACTERISTICS ITEM DESCRIPTIONUNIT CAPABLE OF QUIET OPERATION NO MORE THAN 35 DB 3 FT FM PATIENT; SHALL MEET ISO 9001 STANDARDS FOR LEAKAGE CURRENT; UL APPROVED 3 POWER CORD AND PLUG SUITABLE FOR HOSPITAL ENVIRONMENTS; 110-120 VAC, 60 HZ, 9.5 AMP; 100 VAC, 50/60 HZ, 9.5 AMPS; 220-240 VAC, 50 HZ, 4.5 AMPS DEVICE RATINGS; RUGGED CONSTUCTION TO HANDLE NORMAL MILITARY FIELD SETTING HAZARDS; CARRYING CASE OR CAN FIT INTO FIELD CHEST NO. 6 W/FOAM INSERTS; SUPPLIED W/ALL ANCILLARY ITEMS AND ACCESSORIES TO INCLUDE AS MINIMUM: HOSE W/3 IN. INLET, 2.26 IN. OUTLET, SWIVEL, 6.5 FT LG; HIGH EFFICIENCY 0.2 MICRON FILTER COMPATIBLE W/UNIT-2; BIOMED TESTING AND REPAIR EQUIP/TOOLS NECESSARY FOR FIELD MAINTENANCE/REPAIR; AND INCLUDES CONSUMABLE PATIENT USE ITEMS: UPPER BODY BLANKET; FULL BODY SURGICAL BLANKET; 2 OPERATOR AND 2 SERVICE MANUALS
SPECIAL FEATURESAUTOMATIC PATIENT CONTROL SYSTEM DESIGNED TO REGULATE PATIENT TEMPERATURE BY THE APPLICATION OF HYPERTHERMIA BLANKETS THROUGH WHICH TEMPERATURE CONTROLLED AIR IS CIRCULATED; 3 MODES: HIGH: OP TEMP: 43 DEG C; MEDIUM: OP TEMP;38 D C; AND LOW: OP TEMP 32 DEG C; ALL +/- 3 DEG; OPERATOR SELECTABLE TEMP DISPLAY FOR EITHER CELSIUS OR FAHRENHEIT; WITH CONTROL PANEL WITH THESE FEATURES: MAIN POWER INDICATOR ILLUMIATES WHEN CONNECTED TO POWER SOURCE; SYSTEM ON/STANDBY; OVERHEAT INDICATOR; A TEMP SELECT BUTTON TO ADJUST TEMP; AUDIBLE ALARM TO INDICATE HOSE DISCONNECT AND TEMP FAILURE TO MEET DESIRED SETTING; SHALL NOT BE AFFECTED BY OPERATION OF OTHER ELECTRICAL SYSTEM. UNIT SHOULD BE NO LARGER THAN 33H, 25W, AND 28L, CM; WEIGH NO MORE THAN 5.2 KG;MUST EASILY ATTACH TO IV POLE OR OPT ROLLING STAND; EASY BLANKET ATTACHMENT AND POSITIONING WITH 3 SNAP FIT HOSE SWIVELS AT 3 POINTS; CLAMP THAT SECURES UNIT TO IV POLE
PART NAME ASSIGNED BY CONTROLLING AGENCYTHERMOREGULATOR PATIENT SUITABLE F/HOSPITAL ENVIRONMENT